Revolution Temperature Control

RPA-temp_0.jpg

 

The Revolution Temperature Control option can heat samples from room temperature up to 250 degrees Celsius while running flow tests. Samples can be heated moving constantly, moving intermittently or not moving. Tests can be performed before heating and at temperature intervals.

 

Example Test Data

 

Sample 26C 110C 135C
Polymer Avalanche Angle 51.8 deg Avalanche Angle 60.1 deg Avalanche Angle 68.2 deg
  Density 0.436 g/cm3 Density 0.379 g/cm3 Density 0.372 g/cm3
Polymer Annealed Avalanche Angle 50.5 deg Avalanche Angle 58.0 deg Avalanche Angle 63.3 deg
  Density 0.438 g/cm3 Density 0.390 g/cm3 Density 0.383 g/cm3

 

A polymer was tested at 26C, 110C, and 135C to determine flow changes with temperature between annealed and non-annealed. Increases in temperature caused the powders to flow more poorly and this resulted in a lower powder bed density. In this case the annealed material showed more flow temperature resistance than the non-annealed material.